DETAILS OF MFR HIGH VALUE ITEMS

HOME

Go To Equipment

SL / No Name of Item Name & address of the Firms Date of AT and date of recd by this unit Specification Remarks
1. Oxygen concentrator M/S M.G.M ASSOCIATES, NEW DELHI Procured by BSF 29.08.2007 08.12.2007

Dimension : 28" h x 16" w x 14" d Weight : 52 lbs Delivery Rate : ½ - 5 liters (1/2 Liter increments) Oxygen % : 93% @ 1-5 liters (=/-3%) Electric Requirement : 115 v, 60 Hz Power Consumption : 400 watts average. 1. Pressure "“ relief value and thermal protection on the compressor 2. Filtration for longer bed life and better oxygen purely. 3. Increased flow rates, decreased operating temperatures and longer life with less cup seal wear. 4. Double insulated unit. two prong plug. 5. Humidifier port helps prevent humidifier bottle and connection breakage. 6. The opening is wide to make it easy failure alarm. 7. High/ Low pressure alarms and power failure alarm. 8. Long cannula while maintaining an accurate flow setting (maximum 50 feet) 9. Audible and Visual sensing device for low pressure, power fail and power on.

 
2. ECG Machine with Analyzer    

Power Supply AC Mains/Built in Rechargeable battery operation Mains : 110/220 v ± 10% 48-60 Hz Rechargeable battery : 2 x 6.3 V; 2.6 Ah; Power consumption : 22 VA Fuses : Battery fuse-1.25 A;(220 V) or 0.44(110W) Recording Paper : Thermo sensitive Paper width : 50 mm (recording: 40 mm) Recording speeds : 25 and 50 mm /sec (switch selection) Recording length on RC Battery : 100 m at 25 mm/Sec or 200 m at 50 mm/sec for typical one battery charge. LEAD SELECTION Leads : Electronic 12 leads selection Selector position : Test, I,II,III, aVR, aVL, aVF,VFREQUENCY RESPONSE Frequency response : 0.3 Hz to 120 Hz (AHA) without Filter 0.3 to 100 Hz (IEC) with filter Time constant : 1.5 seconds Filter : Muscle-tremor filter SENSITIVITY Sensitivity : 1.5 mm/mV or greater, continuously variable from 0 to 15 mm/mv with marking for 10 mm/mV calibration. Calibration : 1 mV push-button, generated by electronic circuitry. Damping : 5-10 % overshoot Input impedance : 10 M ohms CMMR : Batter than 80 Db GENERAL: Dimensions : 297 x 210 x 80 mm Weight : 4.2 kg without batteries

 
3. Pulse Oxymeter M/S SAMPARK MEDICARE, B-4, 92A KESHAV PURAM, DELHI-110035 Procured by BSF 20.06.2007 08.12.2007

MEASUREMENT RANGE: SPO2 2 1-100 % P ulse Rate 20-300 beats per minute (bpm) Accuracy SpO2 Adults 70-100 % + 2 digits 0-69% specified Neonates 70-100 % + 3 digits 0-369 % unspecified. Pulse Rate 0-250 bpm + 3 bpm sses 68% of the population. Power Requirements 100-240 VAC 50-60 Hz 15 VA Pulse Rating : 250 A/2A Battery Type- Lead acid with 3 hrs of Battery Backup Operating Temperature Instrument : 10 to 45 º C Sensor within physiologic range for specified accuracy. Weight : 1.3 Kg (max) without pole amount screws Size : 60 (H) x 200 (W) 120 (D) mm % SaO2: Range : 0 % - 100 % Resolution : 1% Accuracy : 50-60 % , Range + 3% , 70-100 % Range + 2% Trends : Pulse Rate-1 hr, 12 hr, 24 hr, SAO2,-1 hr, 12 hr, 24 hr Numeric data display: SPO2 & Pulse Rate (LED +LCD) with backlit LCD display for plesthysmograph with adjustable brightness. Auto gain control system.

 
4. Nebulizer M/S M.G.M ASSOCIATES,2 LOK VIHAR APT, VIKASPURI, NEW DELHI Procured BY BSF 04/06/07 08/12/07

1. Aerosol with piston Electro compressor 2. Nebulising speed : 1 cc in 3 minutes 3. Pressure 1:1 bar 4. Air flow 15 liter/ minute 5. Noise Level : 60 dB 6. Granulo metric size of medication: 1-10 microns by means of multi PIPSER component.

 
5. Portable Anesthesia apparatus kit M/s Manchanda Medicos, 17 A/L Shopping Centre, ShalimarBagh, Delhi- 110088. Procured by CISF AT No.290 dtd 23.03.2009

The machine has to be a complete portable anesthesia machine with all accessories built on a light weight metal frame with foldable built in stand. . Provision for halothane vaporizer with glass bottle. . Flow meter unit for oxygen & Nitrous oxide. . Oxygen & Nitrous Oxide regulators with pin-index system separately for . Brain"Ÿs circuit complete with bag mount corrugated tube minimum 40 inch in length . Re-breathing bags of all sizes. . Face mask padded of all sizes (Pediatrics to adult). . Catheter mounts. . Pediatric circuit with T-Piece arrangements . Facility for emergency oxygen & Nitrous. . "œA" type gas cylinders for oxygen & nitrous oxide. . Nitrous oxide tube 9 inch (230 mm) with fine adjustment control. . Max- connection for B & C type gas cylinder. . Tools kit.s . Carrying case for anesthesia machine with accessories.

 
6. Semi Automatic Blood Analyzer M/s Recorders & Medicare Systems (P) Ltd. 181/15, Phase-1 Industrial Area, Chandgarh-02 PROCURED BY CISF AT NO.297 dtd 31.08.2009 03/12/09

01. Dual mode flow cell and cuvette. 02. Built in voltage stabilizer. 03. Specially designed Cuvette block with control for temperature regulation. 04. Capable of saving 200 tests with QC Graph. 05. Option to enter Patient ID. 06. Monochromatic and Bi-chromatic readings. 07. Selectable temperature: 37 o , 30 o or 25 0 C. + 0.1 o C. 08. Calculation of results either by factor or by standard concentration. 09. Versatile software to calculate the Delta OD on screen. 10. Built in Thermal Printer: high resolution, 320 dots per line with full graphic facility. 11. Automatic sensing of cuvette insertion. 12. Built in real time clock. 13. Operator interface: Touch panel & mouse. 14. Linear measurement range: 0.000 to 3.000 absorbance units (A). 15. Optical measurement. 16. Type of filter: interference. 17. Wave Length: 340,405,510,546,578,630nm & two optional positions 18. Half Bandwidth: 10nm + 2nm. 19. Filter Selection: Automatic. 20. Cuvette/Flow cell: Supports both cuvette and flow cell. 21. Flow cell Sipping Volume: 500 microlitre. 22. Flow Cell Volume: 18 Microlitre. 23. Sipping mode: Automatic by specially designed planetary gear pump. 24. Cuvette Volume: 500 microlitre. 25. Temperature control of Cuvette/flow cell Block: By selectable peltier elements. 26. Light Source: Tungsten Halogen 6 V. 10W with life saving features. 27. Display. 50 Graphics LCD. 28. Analysis Mode: Kinetic defferential, ratio: Absorbance, end point, fixed time. 29. Memory capacity: 64 KB non volatile RAM with battery backup:. Report generation for last 1000 sample stored and obtained by date, ID, both by date & ID. 30. Rs. 232.00 serial port: 9600 baud, 1 start, 8 data, 1 stop. 31. Wattage- 50 Watts or as per the instruments load. 32. Operational Temperature: from 10 0 C to (+) 37 o C. 33. Relative Humidity: UP to 85%. 34. Semi micro Cuvetts : 25 Nos. with each unit. 35. Paper Roll (Thermal): 25Nos with each unit. 36. Variable volume pipette. 2 ul to 20 ul : 01 No. with each unit. 5 ul to 50 ul : 01 No. with each unit 50 ul to 100 ul: 01 No. with each unit 100 ul to 1000 ul : 01 No. with each unit. 37. Compatible UPS for the system.

 